Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
4-6
pubmed:dateCreated
1978-2-18
pubmed:abstractText
A triple human catheter was used for the determination of cardiac output during the acute phase of moycardial infarction in 23 patients. The correlation coefficient between thermodilution and the Fick method was 0.91. Precision of the method assessed by multiple repeated determinations was 4.8 or 5.6%. Quadruplicate determinations allowed to detect variations of cardiac output in the range of 0.465 1/min. Variation between two successive determinations of more than 9.4% can probably be attributed to a technical error.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0008-6312
pubmed:author
pubmed:issnType
Print
pubmed:volume
62
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
291-5
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1977
pubmed:articleTitle
Cardiac output monitoring during the acute phase of myocardial infarction: accuracy and precision of the thermodilution method.
pubmed:publicationType
Journal Article