pubmed:abstractText |
Three major problems which may be encountered with endocardial pacemaker electrodes are a lack of stable position, a chronic increase in stimulation threshold, and a diminishing magnitude of the sensed endocardial signal. These problems are particularly manifest in the atrium. Having previously shown that porous metal surfaces can support stable tissue ingrowth in both bloodstream and soft tissue environments, we set out todetermine the performance of porous-surfaced endocardial pacing electrodes in the atrial position. In two groups of six dogs each, J-shaped atrial leads with Elgiloy electrode tips (2.3 mm. in diameter, 2.3 mm. in length), having either conventional smooth surfaces (control) or porous surfaces (20 to 50 micron particle size) produced by powder metallurgy techniques, were positioned in the right atrial a-pendage. Stimulation thresholds and P-wave amplitude were repeatedly measured until the dogs were put to death 30 w-eks following implantation. The presence or absence of electrode fixation was observed and the atrial tissue reaction was examined grossly and by both light and scanning electron microscopy (SEM). The porous-surfaced electrodes demonstrated superior long-term stimulation thresholds which, at 30 weeks, averaged less then one third of those in the control group. In addition, the porous group showed a small but significant improvement in the amplitude of the sensed P wave. None of the smooth-surfaced electrodes showed fixation, and the tissue reaction consisted of a thick layer of granulation and fibrous tissue on the underlying endocardium, widely separating the electrode from the myocardium. In contrast, all of the porous-surfaced electrodes were fexed to the endocardium by fibrous tissue ingrowth into the surface pores. This tissue fixation of the electrode tip in close proximity to underlying myocytes explains their superior performance.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Aug
|
pubmed:issn |
0022-5223
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
78
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
281-91
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading |
pubmed-meshheading:459536-Animals,
pubmed-meshheading:459536-Dogs,
pubmed-meshheading:459536-Electrodes, Implanted,
pubmed-meshheading:459536-Endocardium,
pubmed-meshheading:459536-Evaluation Studies as Topic,
pubmed-meshheading:459536-Granulation Tissue,
pubmed-meshheading:459536-Heart Atria,
pubmed-meshheading:459536-Heart Catheterization,
pubmed-meshheading:459536-Microscopy, Electron, Scanning,
pubmed-meshheading:459536-Pacemaker, Artificial,
pubmed-meshheading:459536-Prosthesis Design,
pubmed-meshheading:459536-Technology,
pubmed-meshheading:459536-Time Factors
|
pubmed:year |
1979
|
pubmed:articleTitle |
The porous-surfaced electrode: a new concept in pacemaker lead design.
|