Evaluation of linearity, carry-over, precision, and accuracy of a Coulter Counter model S-Plus IV prototype showed that they meet the manufacturer's specifications. The instrument also was compared with an earlier model. Correlations of the lymphocyte and granulocyte values in the three-part differential count with manual eye-count and a Technicon Hemalog D90 results were very close to the correlation of eye-count to Hemalog D90 results. The percentage of mononuclear cells showed acceptable correlation to the manual eye-count when monocytes were combined with blasts, eosinophils, and basophils. The medical effectiveness of the three-part differential was determined by comparison with manual eye-counts in 1,084 samples. The false positive and false negative values were 7.38% and 7.84%, respectively. The instrument has acceptable limits of operation. When combined with analysis of histograms by trained personnel, the three-part differential count is capable of screening for abnormalities that require further analysis.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Sep
|
pubmed:issn |
0002-9173
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
84
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
297-306
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading | |
pubmed:year |
1985
|
pubmed:articleTitle |
Evaluation of the Coulter Counter model S-Plus IV.
|
pubmed:publicationType |
Journal Article,
Comparative Study
|