Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
1985-7-25
pubmed:abstractText
The safety and efficacy of transcutaneous temporary pacing were assessed in 21 patients undergoing elective surgical procedures under general anesthesia. Complete capture was achieved in all patients. The pacing threshold was 50 mamp in two patients, 100 mamp in nine, and 200 mamp in ten. The threshold was influenced by electrode position and also, possibly, by age, heart size, and chest size. No adverse effects of pacing were encountered. Transcutaneous pacing is a rapid, safe, and effective means of temporary pacing in the operating room.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Jul
pubmed:issn
0098-7484
pubmed:author
pubmed:issnType
Print
pubmed:day
5
pubmed:volume
254
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
84-6
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1985
pubmed:articleTitle
Transcutaneous temporary pacing in the operating room.
pubmed:publicationType
Journal Article