Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
2
|
pubmed:dateCreated |
1985-5-14
|
pubmed:abstractText |
The MS-2 automated system was compared with Rosco Neosensitabs agar diffusion for the gentamicin susceptibility of a group of 267 organisms isolated from clinical material. With Rosco Neosensitabs agar diffusion agreement with reference minimal inhibitory concentrations was 98.5%, while with the MS-2 system it was 100% if calcium and magnesium-supplemented MS-2 broth was used.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Feb
|
pubmed:issn |
0021-8847
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
58
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
235-8
|
pubmed:dateRevised |
2000-12-18
|
pubmed:meshHeading | |
pubmed:year |
1985
|
pubmed:articleTitle |
A note on gentamicin susceptibility testing: the validity of the MS-2 system and Rosco Neosensitabs agar diffusion.
|
pubmed:publicationType |
Journal Article
|