Based on a double-headed rotating uncollimated scintillation camera system, a positron imaging device was developed. After a rotating data acquisition in coincidence mode, 16 transverse section images are reconstructed by back projection. To obtain a uniform response, a limited angle reconstruction option is incorporated in this process. After correction for the system response by a three-dimensional deconvolution technique, the 16 transverse section images are stored on disk as a standard patient study for further analysis. The system can also be operated in a stationary mode. In this mode longitudinal tomographic images are obtained. Return to single photon scintigraphy is possible by remounting the collimators and by switching off the coincidence electronics.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Dec
|
pubmed:issn |
0161-5505
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
26
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1466-71
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading | |
pubmed:year |
1985
|
pubmed:articleTitle |
A rotating double-headed positron camera.
|