A hygroscopic condenser humidifier was developed for neonates on mechanical ventilation and was evaluated by laboratory tests and clinically. Humidification provided by the unit was measured in the 10- to 50-ml tidal-volume range at ambient temperatures of 24 degrees C and 38 degrees C. The effect of a leaking patient connection on device performance was investigated. Leakage rates were measured routinely in a neonatal ICU and surgery to determine the clinical significance. In the entire tidal volume and temperature range, the unit provided an inspiratory water content in excess of 30 g/m3 when the leak fraction (volume leaked/volume delivered at Y-piece) was less than 15%. This was found in three out of four cases. In about one out of ten cases, the leak exceeded 30%, which invariably led to corrective action, such as repositioning or changing the endotracheal tube. However, even at a 30% leak, a water content of about 26 g/m3 was still available for humidifying the inspired gas, which corresponds to normal physiologic conditions found in the trachea for nasal breathing of room air.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jan
|
pubmed:issn |
0090-3493
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
15
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
51-4
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading | |
pubmed:year |
1987
|
pubmed:articleTitle |
Neonatal hygroscopic condenser humidifier.
|
pubmed:publicationType |
Journal Article
|