pubmed:abstractText
It has been assumed that the molecular weight (MW) cut-off of a newly fabricated polysulfone capillary dialyzer (F60, Fresenius, FRG) is similar to that of the human glomerulus. We recently tested the device in vivo and found this not to be so, based on the device's ability to eliminate substances of a MW of 10,000 to 60,000 daltons. One of the reasons for this discrepancy was found to be the influence of secondary membrane formation on solute permeability. Endogenous marker substances of a defined MW (beta 2-microglobulin, myoglobin, RBP, alpha 1-microglobulin, acid alpha 1-glycoprotein, alpha 1-antitrypsin, prealbumin, and albumin were measured by laser nephelometry or radioimmune assay; sieving coefficients (SC) and protein eliminations were calculated for each low MW protein.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0085-2538
pubmed:author
pubmed:issnType
Print
pubmed:volume
30
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
429-32
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1986
pubmed:articleTitle
Permeability and secondary membrane formation of a high flux polysulfone hemofilter.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't