We describe a laboratory assessment of three new monitors of blood glucose concentrations: the Boehringer "Accu-Chek II" (B), the Ames "Glucometer II" (A), and the Lifescan "Glucoscan 2000" (L). Inherent imprecision (CV) of each monitor was less than 2%. Maximum difference between individual monitors of the same type was less than or equal to 0.5 mmol/L. The volume of blood applied to the test strips is not critical, but duration of blood incubation or color development should be precise. Two types of test strips retained sufficient color 48 h after development to allow checking of the original measurement, and would be suitable as quality-control "spot" checks. Correlation coefficients for results for whole-blood glucose vs those for serum glucose (measured with the Beckman ASTRA-8) were: 0.992 (B), 0.967 (A), and 0.988 (L). Bias plots of these data showed positive bias for A (0.45 mmol/L) and L (0.17 mmol/L) in relation to serum-glucose measurements, but a negative bias of 0.32 mmol/L for B. Calibration solutions are not interchangeable. Although these versions of the monitors are probably not analytically superior to earlier models, they are easier to use.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
0009-9147
pubmed:author
pubmed:issnType
Print
pubmed:volume
32
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
2195-200
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1986
pubmed:articleTitle
Laboratory assessment of three new monitors of blood glucose: Accu-Chek II, Glucometer II, and Glucoscan 2000.
pubmed:publicationType
Journal Article, Comparative Study, Research Support, Non-U.S. Gov't