pubmed:abstractText
In the past few years, we have occasionally observed linear bands in the phase images of gated cardiac blood-pool scans along the interventricular septum region among patients with normal septal motions. Our retrospective study investigated the cause of out-of-phase bands of 12 patients. We documented through review of cines, computer analysis of gated cardiac blood-pool scintigraphy data, and correlation with echocardiograms that this phenomenon was artifact introduced by rotational movements of the heart. It is important for nuclear physicians to recognize this rotational artifact on the phase analysis image in order to avoid the erroneous misdiagnosis of wall motion abnormalities of the septum.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0161-5505
pubmed:author
pubmed:issnType
Print
pubmed:volume
28
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1536-9
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1987
pubmed:articleTitle
Rotational artifact in phase imaging of cardiac scans: potential pitfalls in diagnosis.
pubmed:affiliation
Department of Radiology, Boston City Hospital, Boston University School of Medicine, Massachusetts.