This paper describes evaluation and correction of count rate characteristics of POSITOLOGICA II, a multi-slice whole body positron emission tomography system. The present study was performed using three phantoms; a 5 cm inner diameter, water-filled lucite cylinder, a 20 cm inner diameter, water-filled lucite cylinder and a chest phantom. After injection of high activity (about 1.85 GBq (50 mCi] of 13N ammonia into each phantom, rates of true coincidence, random coincidence and single photon detections were measured during decay of the isotope through more than two orders of magnitude of activity. At very high levels of activity, count rate characteristics of the system were saturated and limited to 660 kcps of total coincidence rate, which was the sum of rates in on-time and off-time windows, by the FIFO (first-in first-out) output frequency. Below those levels of activity the relationship between count loss and true coincidence rate was not unique but depended on the phantom configurations, suggesting that count loss correction using the above relationship was inadequate for quantitative study. However, the relationship between count loss and single rate was almost independent of the phantom configurations. Thus in conclusion count loss could be corrected using single rate for POSITOLOGICA II. A practical method of count loss correction was also proposed.
pubmed:language
jpn
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
May
pubmed:issn
0033-8303
pubmed:author
pubmed:issnType
Print
pubmed:volume
36
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
221-6
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1987
pubmed:articleTitle
[Count rate characteristics and count loss correction of Positologica II: a whole body positron emission tomograph].
pubmed:publicationType
Journal Article, English Abstract