Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
2
pubmed:dateCreated
1987-5-27
pubmed:abstractText
The type and the incidence of complications during treatment with high-frequency jet ventilation were evaluated in 10 critically ill patients with acute respiratory failure. HFJV was used for 2 to 34 days for management of bronchopleural fistulae, tracheal rupture, laryngeal trauma or voluminous lung abscesses. The most significant technical problems observed were disconnection or kinking of the jet catheter, hypothermia and CO2 retention. Insufficient humidification could induce severe complications such as viscous bronchial secretions, desiccation of the tracheobronchial mucosa or total obturation of the endotracheal tube. Psychological tolerance of high-frequency jet ventilation was generally satisfactory but the ventilator noise was sometimes hardly tolerated. Patients could develop a psychological dependence to high-frequency jet ventilation, leading to weaning problems. Solutions are suggested to decrease the incidence and severity of the technical and psychological complications.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0342-4642
pubmed:author
pubmed:issnType
Print
pubmed:volume
13
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
96-9
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1987
pubmed:articleTitle
Technical and psychological complications of high-frequency jet ventilation.
pubmed:publicationType
Journal Article