Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
1977-7-18
pubmed:abstractText
An improved catheter infusion system is needed for esophageal intraluminal manometry. Using conventional syringe-pump infusion systems undesirably rapid infusion rates of 6 ml per min or more are often needed to achieve accurate recording of esophageal peristaltic pressure. These rapid base line infusion rates are necessitated by the high compliance of syringe-pump systems which causes substantial reduction in the infusion rate during dynamic pressure transients. In this study we tested a hydraulic-capillary infusion system designed to have low compliance. This minimally compliant system yields accurate recording of esophageal peristaltic pressure at infusion rates of 0.6 ml per min or less. We believe that the hydraulic-capillary infusion system is a useful tool for performing both clinical and investigative studies of esophageal motor function.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Jul
pubmed:issn
0016-5085
pubmed:author
pubmed:issnType
Print
pubmed:volume
73
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
23-7
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1977
pubmed:articleTitle
Improved infusion system for intraluminal esophageal manometry.
pubmed:publicationType
Journal Article, Clinical Trial, Research Support, U.S. Gov't, P.H.S.