pubmed:abstractText
Reflectometric measurement of urine test-strips, using the Urotron RL9, was employed to quantify urinary glucose and protein. Although the test-strips (Combur-9-Test) are designed for qualitative use, the reflectance intensities determined by the Urotron RL9 are applicable to quantitative analysis. Reflectance quantitation and conventional colorimetric methods were closely correlated for glucose (r = 0.953) and protein (r = 0.906). Intra-assay coefficients of variation of reflectance value were less than 5%. The negative influence of ascorbic acid on glucose determination was defined quantitatively. A significant positive interference from hemoglobin on protein determination was also demonstrated, but effectively eliminated after compensating for urine color.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0009-9120
pubmed:author
pubmed:issnType
Print
pubmed:volume
21
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
271-5
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1988
pubmed:articleTitle
Quantification of urinary glucose and protein with test-strips through reflectometric analysis.
pubmed:affiliation
Department of Laboratory Medicine, Kumamoto University Medical School, Japan.
pubmed:publicationType