Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
4
|
pubmed:dateCreated |
1989-1-25
|
pubmed:abstractText |
A fluorescence polarization immunoassay (FPIA) procedure for measuring isepamicin, a new aminoglycoside antibiotic, was evaluated. The range of the assay studied was from 0.3-50 micrograms/ml and the calibration curve can be stored for at least 32 days. There was no interference from hemoglobin (up to 400 mg/dl) and bilirubin (up to 20 mg/dl). Within-run precision on each 20 replicates at 5, 15, and 25 micrograms/ml yielded a coefficient of variation (CV) of 1.44-2.52%; between-run precision on analyses at 5, 15, and 25 micrograms/ml over five periods yielded a CV of 1.26-1.59%. Cross-reactivities of the assay system with tobramycin and amikacin were null, but that with gentamicin, of which isepamicin is a derivative, was approximately 14-29%. Data obtained by microbiological assay and FPIA and by high-performance liquid chromatography and FPIA correlated significantly, with coefficients of correlation being 0.980 (n = 70) and 0.965 (n = 39), respectively. The FPIA offers a rapid, efficient, and accurate system for therapeutic monitoring of isepamicin plasma levels.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical |
http://linkedlifedata.com/resource/pubmed/chemical/Amikacin,
http://linkedlifedata.com/resource/pubmed/chemical/Bilirubin,
http://linkedlifedata.com/resource/pubmed/chemical/Gentamicins,
http://linkedlifedata.com/resource/pubmed/chemical/Tobramycin,
http://linkedlifedata.com/resource/pubmed/chemical/isepamicin
|
pubmed:status |
MEDLINE
|
pubmed:issn |
0163-4356
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
10
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
459-62
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:3201531-Adult,
pubmed-meshheading:3201531-Amikacin,
pubmed-meshheading:3201531-Bilirubin,
pubmed-meshheading:3201531-Biological Assay,
pubmed-meshheading:3201531-Chromatography, High Pressure Liquid,
pubmed-meshheading:3201531-Drug Interactions,
pubmed-meshheading:3201531-Gentamicins,
pubmed-meshheading:3201531-Hemolysis,
pubmed-meshheading:3201531-Humans,
pubmed-meshheading:3201531-Immunoassay,
pubmed-meshheading:3201531-Male,
pubmed-meshheading:3201531-Reproducibility of Results,
pubmed-meshheading:3201531-Tobramycin
|
pubmed:year |
1988
|
pubmed:articleTitle |
Evaluation of a fluorescence polarization immunoassay procedure for quantitation of isepamicin, a new aminoglycoside antibiotic.
|
pubmed:affiliation |
Department of Pharmacology, Hamamatsu University School of Medicine, Japan.
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Research Support, Non-U.S. Gov't
|