The purpose of this investigation was to test the reliability of the American Optical Non-Contact Tonometer Mark II (NCT II) using the Goldmann Applanation Tonometer (GAT) as the validating instrument. The sample contained 102 consecutive patients from our University Eye Clinic, of whom one-half had 4 NCT II measurements first, followed by 4 GAT measurements; the other one-half had 4 GAT measurements first, followed by 4 NCT II measurements. No significant change in intraocular pressure (IOP) was noted over the measurement sequence with either instrument. There was no significant difference between paired NCT II and GAT readings when the NCT II was used first; however, a highly significant difference between paired readings was obtained when the GAT was used first, indicating that the GAT measurement produced a delayed reduction in the IOP. This effect did not occur with the NCT II. Although the NCT II is shown to have a good overall reliability when compared to the GAT in both protocols, the agreement between any two tonometers may be influenced greatly by the very process of taking a measurement and by the dynamic nature of the IOP.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0093-7002
pubmed:author
pubmed:issnType
Print
pubmed:volume
65
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
751-6
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1988
pubmed:articleTitle
Clinical evaluation of the Non-Contact Tonometer Mark II.
pubmed:affiliation
Department of Optometry, University of Wales Institute of Science and Technology, Cardiff, United Kingdom.
pubmed:publicationType
Journal Article