Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
4
pubmed:dateCreated
1988-11-3
pubmed:abstractText
Thirty-one patients with severe respiratory failure who were failing volume controlled conventional ratio ventilation were placed on pressure controlled inverse ratio ventilation (PC-IRV) for a total of 4,426 patient-hours. The PC-IRV resulted in a reduction of minute ventilation from 22 +/- 1.0 L/min (mean +/- SEM) to 15 +/- 0.7 L/min. Peak inspiratory pressure (PIP) was reduced from 66 +/- 2.3 cm H2O to 46 +/- 1.6 cm H2O and positive end expiratory pressures (PEEP) from 15 +/- 1.0 cm H2O to 2.5 +/- 0.5 cm H2O. Mean airway pressure increased from 30 +/- 1.7 cm H2O to 35 +/- 1.7 cm H2O. Oxygenation (PaO2) improved from 69 +/- 4.0 mm Hg to 80 +/- 4.5 mm Hg. The PaCO2 and arterial pH were not significantly changed. There were no significant changes in mean hemodynamic pressures. A lung compromise index (FIO2.PIP.10/PaO2) retrospectively distinguished between successful and unsuccessful PC-IRV episodes. These data suggest that PC-IRV can be successfully and safely implemented in critically ill patients with severe respiratory failure over prolonged periods of time resulting in significant improvement in oxygenation at lower minute volume, peak airway pressure and PEEP requirements.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0012-3692
pubmed:author
pubmed:issnType
Print
pubmed:volume
94
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
755-62
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1988
pubmed:articleTitle
Pressure controlled inverse ratio ventilation in severe adult respiratory failure.
pubmed:affiliation
Division of Pulmonary and Critical Care Medicine, University of California, Davis Medical Center, Sacramento.
pubmed:publicationType
Journal Article