The method described in this paper permits determination of amplification, band-pass properties, and electrode impedance in electrophysiological experimental setups. The battery powered test-signal generator drives square-wave current pulses through a 2 ohm resistance, which acts as a source of 100 muV p-p pulses. The 2 ohm resistor is wired in series with one of the electrode leads and remains there permanently. For calibration of the above-mentioned properties, one need only turn the generator on, without changing the measurement setup. By observing the shape of the square waves after passing through the system, it is possible to judge the band-pass characteristics of the entire system. Electrode impedance may be measured by putting a (variable) resistor in parallel with the preamplifier input. The test signal current is of the same order of magnitude as currents from biological signal sources. This method makes working with electrophysiological measurement systems which are used in electroretinography, electrooculography, and recording of visually-evoked cortical potentials in ophthalmology much easier.
pubmed:language
ger
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
0065-6100
pubmed:author
pubmed:issnType
Print
pubmed:day
31
pubmed:volume
205
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
33-9
pubmed:dateRevised
2009-11-11
pubmed:meshHeading
pubmed:year
1977
pubmed:articleTitle
[A new method for calibration of electrophysiological recording systems (author's transl)].
pubmed:publicationType
Journal Article, English Abstract