Five studies were conducted to examine the degree of variability to be expected during the use of the non-imaging nuclear probe (BIOS Inc.) under a variety of clinical conditions. Comparison of the ejection fraction (EF) readings between the nuclear probe and a gamma camera showed good agreement, with the nuclear probe tending to underestimate lower, and overestimate higher camera EF values [mean (S.D.) difference, 0.84% (6.06)]. A comparison of two nuclear probes showed a small mean (S.D.) difference of EF readings of 0.063% (2.26). EF readings obtained in normal subjects 6 weeks apart were reproducible and differed by a mean (S.D.) of 0.23% (4.42). The administration of placebo to 10 normal subjects followed by sequential measurements for 4 h produced EF changes large enough to mimic a clinical effect, the largest hourly change observed being 5.4%, indicating the need for strict placebo control in interventional experiments. Data on four patients with heart failure showed small non-significant EF changes in the 1 h after placebo administration but a wide intra-subject range of ejection time and time to peak filling measurements. This highlights the problem of accurate, reproducible cursor placement in such patients. The nuclear probe is a portable, low cost instrument which produces accurate EF measurements when compared with the gamma camera.
|
pubmed:commentsCorrections | |
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Oct
|
pubmed:issn |
0143-3636
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
7
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
717-28
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading |
pubmed-meshheading:3027633-Adolescent,
pubmed-meshheading:3027633-Adult,
pubmed-meshheading:3027633-Aged,
pubmed-meshheading:3027633-Aged, 80 and over,
pubmed-meshheading:3027633-Heart,
pubmed-meshheading:3027633-Heart Diseases,
pubmed-meshheading:3027633-Heart Function Tests,
pubmed-meshheading:3027633-Humans,
pubmed-meshheading:3027633-Middle Aged,
pubmed-meshheading:3027633-Sodium Pertechnetate Tc 99m,
pubmed-meshheading:3027633-Stroke Volume
|
pubmed:year |
1986
|
pubmed:articleTitle |
A comprehensive clinical validation of the nuclear stethoscope.
|
pubmed:publicationType |
Journal Article
|