pubmed:abstractText
The biocompatibility of heparin surface modified poly(methyl methacrylate) intraocular lenses (IOLs) was evaluated in two experiments following implantation in the anterior and posterior eye chambers of adult cynomolgus monkeys. Throughout the study, large inflammatory cells and prominent pigment deposits were seen on the unmodified lenses, whereas the heparin surface modified IOLs remained almost free of precipitates. Similarly, fewer posterior synechias were observed in eyes implanted with surface modified IOLs in the posterior chamber than in eyes implanted with control lenses. Histopathological examination of enucleated eyes confirmed the clinical findings. These experiments strongly support the idea that surface modification with heparin is a useful way to reduce clinical complications following cataract surgery with IOL implantation.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0886-3350
pubmed:author
pubmed:issnType
Print
pubmed:volume
15
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
485-90
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1989
pubmed:articleTitle
Heparin surface modified intraocular lenses implanted in the monkey eye.