Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
5
|
pubmed:dateCreated |
1989-10-19
|
pubmed:abstractText |
The effect of the boundary conditions between trabecular bone specimens and the test columns of the testing machine was studied together with the effect of side-constraint on the mechanical behaviour of trabecular bone during axial compression. Cylindrical specimens taken from the upper tibial epiphysis of autopsy knees were tested non-destructively by cyclic compression to 0.8% strain under different conditions. Fixation of the specimens to the test columns by a thin layer of bone cement increased the stiffness by 40% and reduced the energy dissipation to 67% of those measured under unconstrained conditions (p less than 0.001). The thin cement layer alone increased the stiffness 19% and reduced energy dissipation to 86% (n.s.). When the machine was equipped with polished steel columns coated by a film of low-viscous oil, both the stiffness and the energy dissipation were reduced to 93% of those measured under standard conditions (p less than 0.005). Trabecular bone specimens tested side-constrained by the surrounding trabecular bone (in situ) showed a 19% larger stiffness than that measured during later testing of the corresponding machined specimens (p less than 0.005) whereas the energy dissipation was not altered significantly. The same specimens showed a 22% increase of stiffness and a 68% increase of energy dissipation when they were side-constrained by a closely fitting steel cylinder (p less than 0.005).
|
pubmed:commentsCorrections | |
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:issn |
0021-9290
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
22
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
485-90
|
pubmed:dateRevised |
2009-11-11
|
pubmed:meshHeading |
pubmed-meshheading:2777823-Adult,
pubmed-meshheading:2777823-Aged,
pubmed-meshheading:2777823-Aged, 80 and over,
pubmed-meshheading:2777823-Biomechanics,
pubmed-meshheading:2777823-Bone and Bones,
pubmed-meshheading:2777823-Elasticity,
pubmed-meshheading:2777823-Humans,
pubmed-meshheading:2777823-Middle Aged,
pubmed-meshheading:2777823-Reference Values,
pubmed-meshheading:2777823-Research Design,
pubmed-meshheading:2777823-Tensile Strength,
pubmed-meshheading:2777823-Tibia
|
pubmed:year |
1989
|
pubmed:articleTitle |
The effect of constraint on the mechanical behaviour of trabecular bone specimens.
|
pubmed:affiliation |
Biomechanics Laboratory. Orthopaedic Hospital, University of Aarhus, Denmark.
|
pubmed:publicationType |
Journal Article,
Research Support, Non-U.S. Gov't
|