Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
4
pubmed:dateCreated
1989-10-5
pubmed:abstractText
An x-ray fluorescence system which utilizes polarized radiation to measure lead in vivo in human subjects is described. The minimum detection limit is approximately 6.4 ppm wet weight lead in the cortex of the tibia with 4 mm of overlying soft tissue. This appears to be adequate for assessing lead stores in lead-toxic preschool children. The measurement requires 16.5 min and is associated with an effective equivalent whole body dose to the subject of 2.5 muSv. The system, its calibration and its validation are described herein.
pubmed:grant
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
0094-2405
pubmed:author
pubmed:issnType
Print
pubmed:volume
16
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
521-8
pubmed:dateRevised
2007-11-14
pubmed:meshHeading
pubmed:articleTitle
In vivo measurement of cortical bone lead using polarized x rays.
pubmed:affiliation
Radiation Oncology Department, University Hospital, State University of New York, Stony Brook 11794-7028.
pubmed:publicationType
Journal Article, Research Support, U.S. Gov't, P.H.S.