Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
11
pubmed:dateCreated
1990-5-18
pubmed:abstractText
A simple and sensitive high-performance liquid chromatographic method with fluorescence detection for the determination of nineteen 21-hydroxycorticosteroids is described. The corticosteroids are oxidized by cupric acetate to form the corresponding glyoxal derivatives. The derivatives are converted into fluorescent quinoxalines by reaction with 1,2-diamino-4,5-methylenedioxybenzene, a fluorogenic reagent for alpha-dicarbonyl compounds. The quinoxalines are separated within 70 min on a reversed-phase column (TSK gel ODS-120T) by stepwise elution with mixtures of methanol, acetonitrile, and 1.0 M ammonium acetate. The detection limits are 0.14-29.4 pmol at a signal-to-noise ratio of 3 in a 50-microliter injection volume. This sensitivity permits precise determination of hydrocortisone, cortisone, corticosterone, and their tetrahydro derivatives in 500 microliters of normal human urine.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0009-2363
pubmed:author
pubmed:issnType
Print
pubmed:volume
37
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
3022-5
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1989
pubmed:articleTitle
Determination of 21-hydroxycorticosteroids in human urine by high-performance liquid chromatography with fluorescence detection.
pubmed:publicationType