Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
3
pubmed:dateCreated
1981-9-22
pubmed:abstractText
We describe a sensitive and precise "high-pressure" liquid-chromatographic method for determining acetaminophen in serum. The 2-acetaminophenol and 3-acetaminophenol structural isomers of acetaminophen are used as internal standards. The method, which involves solvent extraction and adsorption chromatography on silica, provides excellent sensitivity, accuracy, and selectivity. The standard curve is linear over the range of acetaminophen concentrations of 0.5 to 300 mg/L, which makes the method useful for both pharmacokinetic studies and overdose monitoring. Analytical recovery is 97% for acetaminophen concentrations ranging from 5 to 300 mg/L. Many commonly used drugs were tested and found not to interfere. The procedure has been successfully adapted as a microscale method requiring only 50 microL of sample. The microscale method is particularly useful for pediatric and neonatal patients for whom sample size is a major concern.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Mar
pubmed:issn
0009-9147
pubmed:author
pubmed:issnType
Print
pubmed:volume
25
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
409-12
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1979
pubmed:articleTitle
Liquid-chromatographic determination of acetaminophen in serum.
pubmed:publicationType
Journal Article, Comparative Study