pubmed:abstractText
A new molded polypropylene, diamond-like carbon (DLC)-coated mixing cell has been developed for use in the batch microcalorimeter. Reagent volume can be varied from 25 microliters to 100 microliters. A 10 microcalorie reaction heat can be measured to 5%. Repeat reactions can be done as often as every 10 min for a fast reaction. Reactions can be started within 1 h or less after loading. A pre-equilibrator and a temperature-controlled syringe drive unit permit solutions to be stored at 4 degrees C while being run at any temperature from -20 degrees C to 40 degrees C. The kinetics and enthalpy of reaction of NAD-NADase have been measured. delta H is about 21 kcal/mol endothermic.