A microprocessor-based timer has been developed for routine and research use in cardiac electrophysiologic studies. The cycle length (A-A) and the conduction times through the right atrium (A(HRA)-A(HIS], the AV node (A-H) and the His-Purkinje system (H-V) can be automatically measured beat-to-beat in real-time both during sinus rhythm and during atrial pacing. The design has been refined during three years of use in over 80 clinical studies. A comparison between manual and automatic measurements has been executed for randomly chosen portions of 10 studies. With the manual measurements as reference procedures the sensitivities of the automatic measurements of the A, H and V waves were 0.98, 0.88 and 0.92, respectively, while the specificities were 0.93, 0.92 and 0.93, respectively. For the true positive measurements the correlation coefficients of the intervals were r(A-A) = 0.9998, r(A-H) = 0.987 and r(H-V) = 0.988. The statistical analysis of the differences between manual and automatic readings for the intervals A-A, A-H and H-V yielded mean values of -1 ms, -0.8 ms and 0.6 ms and standard deviations of 3 ms, 4 ms and 4 ms which are smaller than the accuracy of the common manual measurements.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
11
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
33-46
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1988
pubmed:articleTitle
On-line interval measurement during invasive cardiac electrophysiologic testing.
pubmed:affiliation
Dipartimento di Fisica, Universitá di Trento, Italy.