The purpose of this study was to evaluate the use of exercise testing in identifying abnormalities of pacemaker function and in confirming set parameters not apparent at rest in children with implanted atrial synchronous physiologic pacemakers. Maximal exercise tests were performed on 24 children (15 boys, 9 girls) from 4.5 to 18 years of age (median = 15) with physiologic pacemakers. The lower rate limit was observed before or following testing in 19 of 24 cases. In each case this correlated with the set lower rate limit. The upper rate limit was reached in 10 of 24 cases and was found to be lower than that programmed in one case in which a long atrial refractory period had limited the upper rate limit. Six children reached the maximum upper rate limit to which their pacemaker could be programmed. No abnormalities of atrial capture or ventricular capture occurred during exercise testing. Ventricular sensing was normal in each case. Atrial sensing was observed to be normal in 15 of the 24 cases. Two patients had decreased atrial sensing with exercise. Reversion to the "noise rate" due to myopotential inhibition was found in seven other cases. Subsequent tests on two of these children showed normal sensing.(ABSTRACT TRUNCATED AT 250 WORDS)
pubmed:grant
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
8
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
656-60
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1985
pubmed:articleTitle
The use of exercise testing in children to evaluate abnormalities of pacemaker function not apparent at rest.
pubmed:publicationType
Journal Article, Research Support, U.S. Gov't, P.H.S.