pubmed:abstractText
The suitability of a test for standard specification testing depends inter alia on the ability to repeat the tests in a reproducible manner at a number of test centres. This work involved the investigation of three tests currently proposed as standard specification tests for dental materials. It was found that one test (compressive strength of cements) is inappropriate for inclusion in standards due to an unacceptable variation in test results between test centres. The treatment of results suggested in standards should place less emphasis on the mean value of a relatively small number of test specimens and the use of a simple form of probability theory in which, say, 80 per cent of specimens are required to achieve a certain pass level. In some standards there may be a need to increase the numbers of test specimens significantly in order to achieve a more meaningful and reliable result. The use of Weibull statistics can be adopted as a means of identifying tests which are suitable for inclusion in standard specifications, although it is doubtful that a test at this level of sophistication is required or desirable for inclusion in the standards themselves.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
D
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0300-5712
pubmed:author
pubmed:issnType
Print
pubmed:volume
18
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
90-7
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1990
pubmed:articleTitle
An investigation of test-house variability in the mechanical testing of dental materials and the statistical treatment of results.
pubmed:affiliation
Dental School, University of Newcastle upon Tyne, UK.
pubmed:publicationType
Journal Article, Comparative Study