Ilizarov proposes the use of a special circular fixation device for treatment of bone defects and nonunions or for limb lengthening. Supposition was that the success of this method has to do with the specific mechanical behavior of the device. This behavior is a result of the configuration of the fixation elements. Therefore, a mechanical study of the sensitivity of the circular compression and distraction device (CDD) to configuration parameters was performed. The CDD was found to exhibit a nonlinear stiffness behavior, in particular under axial load. This may be favorable for the induction and tolerance of bone formation. Among the different parameters tested the ring radius was the most important with respect to stiffness. In general, the stiffness of the CDD allows adjustment during postoperative management. In magnitude, it is equal or lower when compared to other fixator types.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Feb
pubmed:issn
0148-0731
pubmed:author
pubmed:issnType
Print
pubmed:volume
112
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
15-21
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1990
pubmed:articleTitle
Stiffness characteristics of the circular Ilizarov device as opposed to conventional external fixators.
pubmed:affiliation
M. E. Mueller-Institute for Biomechanics, University of Bern, Switzerland.
pubmed:publicationType
Journal Article, Comparative Study