Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
6
|
pubmed:dateCreated |
1990-10-18
|
pubmed:abstractText |
In May, 1982, a 49-year-old man underwent mitral valve replacement (MVR) in our hospital with a 31 mm Björk-Shiley prosthesis for mitral regurgitation. He had been doing well until his episode of palpitation and dyspnea of sudden onset, and was transferred to our ICU with severe cardiogenic shock in Aug, 1986. Chest X-ray film revealed pulmonary edema and breakage of the valve with migration of the disc and the minor strut of the prosthesis. He was operated upon 5 hours after the onset of his complaints. The minor strut was removed from the left upper pulmonary vein and mitral valve re-replacement was done with a 29 mm Björk-Shiley Monostrut valve. The disc which had dislocated into the abdominal aorta was also recovered on the twenty-third post operative day. His postoperative course was uneventful. Immediate diagnosis and subsequent re-operation is absolute indication for rescue from acute cardiac failure due to mechanical failure of any prosthetic valve.
|
pubmed:language |
jpn
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jun
|
pubmed:issn |
0369-4739
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
38
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1049-52
|
pubmed:dateRevised |
2011-7-27
|
pubmed:meshHeading | |
pubmed:year |
1990
|
pubmed:articleTitle |
[Minor strut fracture of the Björk-Shiley mitral valve].
|
pubmed:affiliation |
Department of Cardiovascular Surgery, Heart Institute Hyogo Kenritsu Amagasaki Hospital, Amagasaki, Japan.
|
pubmed:publicationType |
Journal Article,
English Abstract,
Review,
Case Reports
|