We conducted an European multicentre trial to assess the performance of the new Boehringer Mannheim/Hitachi 717 analysis system. The photometer response was linear up to an absorbance of 2.8. The maximal CV of photometric imprecision was 0.5% for the wavelength pair 340/405 nm within the absorbance range 0.9 to 2.4. For the 13 analytes in our study, mean within-run imprecision was less than 2%, and mean between-day imprecision less than 2.5%. The results obtained with the Hitachi 717 instrument correlated closely with those of comparison instruments. Linearity for the various tests was high and exceeded the manufacturer's claims. No drift was detected during an 8-hour work period; carry over could not be detected under the chosen experimental conditions. The new instrument was readily accepted by the evaluators because of its ease of handling and simple daily maintenance.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0340-076X
pubmed:author
pubmed:issnType
Print
pubmed:volume
28
pubmed:owner
NLM
pubmed:authorsComplete
N
pubmed:pagination
261-9
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1990
pubmed:articleTitle
Multicentre evaluation of the Boehringer Mannheim/Hitachi 717 analysis system.
pubmed:affiliation
Radboud Ziekenhuis, Nijmegen, The Netherlands.
pubmed:publicationType
Journal Article, Clinical Trial, Comparative Study, Multicenter Study