A recently developed chemiluminescence immunoassay system (LIA-mat) for triiodothyronine and thyroxine, set up by Byk-Sangtec Diagnostica (Dietzenbach, Germany), has been evaluated and compared with radioimmunoassays and with a chemiluminescence enhanced enzyme immunoassay (Amerlite), using control materials circulated in a national interlaboratory quality control, as well as patient sera. The LIA-mat assays are competitive methods which use coated monoclonal antibodies and triiodothyronine- or thyroxine-ABEI (aminobutylethylisoluminol) conjugate as tracers. The working range of LIA-mat T3 (computed from the within-assay precision profile) extended from 1.4 to 12.3 nmol/l; the between-assay precision was 8.1 - 19.3 CV%. Regression analysis of the LIA-mat T3 results (y) against the consensus means (x) of the participants in the national interlaboratory survey yielded: y = -0.14 + 1.05 x, r = 0.95. The working range of LIA-mat T4 extended from 33 to 515 nmol/l; the between-assay precision was 5.4 - 9.2 CV%. An excellent agreement was found between LIA-mat T4 results (y) and the consensus means (x) of the laboratories participating in the national interlaboratory survey (y = 3.79 + 1.02 x, r = 0.98).
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Apr
|
pubmed:issn |
0340-076X
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
28
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
193-7
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading |
pubmed-meshheading:2193102-Dose-Response Relationship, Drug,
pubmed-meshheading:2193102-Evaluation Studies as Topic,
pubmed-meshheading:2193102-Humans,
pubmed-meshheading:2193102-Immunoenzyme Techniques,
pubmed-meshheading:2193102-Luminescent Measurements,
pubmed-meshheading:2193102-Methods,
pubmed-meshheading:2193102-Radioimmunoassay,
pubmed-meshheading:2193102-Reference Standards,
pubmed-meshheading:2193102-Regression Analysis,
pubmed-meshheading:2193102-Reproducibility of Results,
pubmed-meshheading:2193102-Thyroxine,
pubmed-meshheading:2193102-Triiodothyronine
|
pubmed:year |
1990
|
pubmed:articleTitle |
A new chemiluminescence immunoassay for triiodothyronine and thyroxine: evaluation using quality control sera assayed in an interlaboratory survey.
|