pubmed:abstractText
The possibility of using thermocouples for the artifact-free measurements of skin temperature during millimeter wave exposure was studied. The distributions of the specific absorption rate (SAR) in the human skin were calculated for different orientations of the thermocouple relative to the E-field of exposure. It was shown that, at the parallel orientation of a thermocouple relative to the E-field, SAR significantly increased at the tip of the thermocouple. This can result in an overheating of the thermocouple. At the perpendicular orientation of a thermocouple, the distortions of the SAR were insignificant. The data obtained confirm that the skin temperature can be measured with a thermocouple during exposure under the condition that the thermocouple is located perpendicular to the E-vector of the electromagnetic field. For the accurate determination of SAR from the rate of the initial temperature rise, it is necessary to fit the temperature kinetics measured with the thermocouple to the solution of the bio-heat transfer equation.
pubmed:grant
pubmed:language
rus
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0006-3029
pubmed:author
pubmed:issnType
Print
pubmed:volume
56
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
561-5
pubmed:meshHeading
pubmed:articleTitle
[Problems of using a thermocouple for measurements of skin temperature rise during the exposure to millimeter waves].
pubmed:publicationType
Journal Article, English Abstract, Research Support, N.I.H., Extramural