Source:http://linkedlifedata.com/resource/pubmed/id/21245064
Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions |
umls-concept:C0022023,
umls-concept:C0022341,
umls-concept:C0034603,
umls-concept:C0086562,
umls-concept:C0180392,
umls-concept:C0200637,
umls-concept:C0444706,
umls-concept:C0556642,
umls-concept:C1521970,
umls-concept:C1551432,
umls-concept:C1881830,
umls-concept:C2240038,
umls-concept:C2347880,
umls-concept:C2827424
|
pubmed:issue |
2-4
|
pubmed:dateCreated |
2011-3-2
|
pubmed:abstractText |
Active mobile dosimetry unit (Liulin), passive plastic nuclear track detectors (PNTD) and thermoluminescent detectors (TLD) were exposed in a C290 MeV/n beam at HIMAC-BIO (NIRS, Japan). Two different types of beam configuration were used--monoenergetic beam (MONO) and spread-out Bragg peak (SOBP); the detectors were placed at several depths from the entrance up to the depths behind the Bragg peak. Relative response of TLDs in beams has been studied as a function of the depth, and it was re-proved that it can depend on the linear energy transfer (LET). Liulin measures energy deposition in Si; the spectra of energy deposited in Si can be transformed to the spectra of lineal energy or LET. PNTDs are able to determine the LET of registered particles directly. The limitation of both methods is in the range in which they can determine the LET-Liulin is able to measure perpendicularly incident charged particles up to ?35 keV/µm (in water), PNTD can measure from ?7 to 400 keV/µm, independently of the registration angle. The results from both methods are compared and combined for both beams' configuration, and a good agreement is observed.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Feb
|
pubmed:issn |
1742-3406
|
pubmed:author | |
pubmed:issnType |
Electronic
|
pubmed:volume |
143
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
519-22
|
pubmed:meshHeading |
pubmed-meshheading:21245064-Equipment Design,
pubmed-meshheading:21245064-Equipment Failure Analysis,
pubmed-meshheading:21245064-Heavy Ions,
pubmed-meshheading:21245064-Japan,
pubmed-meshheading:21245064-Linear Energy Transfer,
pubmed-meshheading:21245064-Particle Accelerators,
pubmed-meshheading:21245064-Radiometry
|
pubmed:year |
2011
|
pubmed:articleTitle |
Spectra of linear energy transfer and other dosimetry characteristics as measured in C290 MeV/n MONO and SOBP ion beams at HIMAC-BIO (NIRS, Japan) with different detectors.
|
pubmed:affiliation |
Department of Radiation Dosimetry, Nuclear Physics Institute, Czech Academy of Sciences, Na Truhlá?ce 39/64, 18086 Praha 8, Czech Republic.
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Evaluation Studies
|