pubmed:abstractText
In patients with an implantable cardioverter defibrillator, empirical antitachycardia pacing/burst pacing (ATP) is associated with a significantly decreased rate of appropriate shocks. The use of ATP as first-line therapy in ventricular tachycardia promotes less pain and better quality of life because the number of shocks is reduced. Additionally, battery longevity is substantially increased with this strategy. Based on this, device manufacturers have developed new algorithms to optimize the use of ATP and shocks in patients with an implantable cardioverter defibrillator. The present report describes a case in which the use of one of these new algorithms was associated with a significant delay in tachycardia termination and, consequently, led to syncope.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
1916-7075
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
26
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
e341-3
pubmed:dateRevised
2011-11-1
pubmed:meshHeading
pubmed:year
2010
pubmed:articleTitle
Implantable cardioverter defibrillator: charge saver, not syncope saver!