Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
1
|
pubmed:dateCreated |
1990-2-1
|
pubmed:abstractText |
It is often necessary to estimate the dose of radiation to a fetus from a series of CT scans. To assist in making this process easier and more accurate, we measured fetal doses in an adult anthropomorphic phantom for four CT scanners: Picker 1200, Siemens DRH, GE 9800, and GE 8800. Measurements were made at four kilovoltages (100, 120, 130, and 140 kVp), for 2-, 5-, 8-, and 10-mm thicknesses, for two scanning angles (360 degrees and 398 degrees) and for two patient orientations (prone and supine). The fetal-dose estimates are based on the CT dose index measured by using a pencil ionization chamber at the center position of a 16-cm-diameter cylindrical phantom. Comparison with data from other investigators shows reasonable agreement when phantom differences, X-ray tube output, and equipment calibration are considered.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jan
|
pubmed:issn |
0361-803X
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
154
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
185-90
|
pubmed:dateRevised |
2008-2-15
|
pubmed:meshHeading |
pubmed-meshheading:2104708-Female,
pubmed-meshheading:2104708-Fetus,
pubmed-meshheading:2104708-Humans,
pubmed-meshheading:2104708-Models, Structural,
pubmed-meshheading:2104708-Pregnancy,
pubmed-meshheading:2104708-Prenatal Diagnosis,
pubmed-meshheading:2104708-Radiation Dosage,
pubmed-meshheading:2104708-Tomography, X-Ray Computed
|
pubmed:year |
1990
|
pubmed:articleTitle |
Estimated fetal radiation dose from multislice CT studies.
|