Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
8
pubmed:dateCreated
2010-9-10
pubmed:abstractText
A 46-year-old man presented to our institution with inappropriate implantable cardioverter-defibrillator (ICD) shock delivery. The ICD (single chamber, dual shock coils) was implanted for sustained monomorphic ventricular tachycardia with unstable hemodynamics and underlying systolic left ventricular dysfunction. ICD interrogation revealed recurrent episodes of ICD shock due to noise sensing and increased impedance of right ventricular (RV)-lead. With the impression of lead fracture, ICD lead extraction was performed. The fractured ICD lead was completely removed by traction of locking stylet and counter-traction of polypropylene dilator sheath. A new lead was inserted and the patient was discharged without complications after 2 days. To our knowledge, this is the first report on ICD lead extraction by conventional traction and counter-traction technique in Korea.
pubmed:language
eng
pubmed:journal
pubmed:status
PubMed-not-MEDLINE
pubmed:month
Aug
pubmed:issn
1738-5555
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
40
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
418-20
pubmed:year
2010
pubmed:articleTitle
Implantable cardioverter-defibrillator lead extraction by conventional traction and counter-traction technique.
pubmed:affiliation
Division of Cardiology, Severance Cardiovascular Hospital, Yonsei University Health System, Seoul, Korea.
pubmed:publicationType