pubmed:abstractText
Twelve articulators (Denar D5-A and Mark II) were calibrated using the Denar Field Inspection Gage device. Silicone interocclusal records and the split cast method were used to test the positional accuracy of mounted casts transferred from one articulator to another. The interchangeability was performed by three independent observers. The results indicate that calibrated articulators can be used to transfer casts between articulators without a loss of positional accuracy.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
D
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0022-3913
pubmed:author
pubmed:issnType
Print
pubmed:volume
65
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
845-8
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1991
pubmed:articleTitle
Determining the accuracy of articulator interchangeability.
pubmed:affiliation
Department of Periodontology, State University of New York, School of Dental Medicine, Stony Brook.
pubmed:publicationType
Journal Article