The Vitek 2 Anaerobe and Corynebacterium Identification Card (ANC) was recently evaluated in a multicentre study. In the present work, this system was compared with the BBL Crystal Anaerobe and RapID ANA II panels. These kits were tested using 196 strains of anaerobes that had been previously identified by gas-liquid chromatography. Identification to the species or to the genus level was 75.0%, 81.1% and 70.9% for Crystal, RapID and Vitek, respectively. Vitek ANC failed to provide any identification in 20.4% of the strains, but it had fewer misidentifications than RapID. The confidence factors provided on the results report of each kit were not always correlated with a lower risk of major errors, with the exception of Vitek 2 in which a confidence factor higher than 0.86 excluded the risk of misidentification in more than 87% of isolates. The lower rate of identification by the Vitek and Crystal panels is mostly due the lower ability of these systems to identify the Clostridia. Overall, the three panels are comparable but need improvement to a better accuracy.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Aug
pubmed:issn
1095-8274
pubmed:author
pubmed:copyrightInfo
Copyright (c) 2010 Elsevier Ltd. All rights reserved.
pubmed:issnType
Electronic
pubmed:volume
16
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
355-61
pubmed:meshHeading
pubmed:year
2010
pubmed:articleTitle
Vitek 2 ANC card versus BBL Crystal Anaerobe and RapID ANA II for identification of clinical anaerobic bacteria.
pubmed:affiliation
Department of Microbiology, St-Pierre University Hospital, Brussels, Belgium. laurent_blairon@stpierre-bru.be
pubmed:publicationType
Journal Article, Comparative Study, Evaluation Studies