Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
7
pubmed:dateCreated
2010-7-23
pubmed:abstractText
In the case presented problems with mechanical and manual ventilation of a patient occurred during the operation. Prior to this endotracheal intubation had been performed without difficulty and the respirator had passed all system checks. A leakage in the recently changed carbon dioxide absorber could be detected which had been accidentally dropped and damaged internally.
pubmed:language
ger
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jul
pubmed:issn
1432-055X
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
59
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
652-4
pubmed:meshHeading
pubmed:year
2010
pubmed:articleTitle
[Cleft in carbon dioxide absorber. Intraoperative problems with ventilation due to a leak in the breathing circuit].
pubmed:affiliation
Klinik für Anästhesiologie und Operative Intensivmedizin, Universitätsklinik Köln, Kerpener Str. 62, 50937 Köln, Deutschland. christian.paul@uk-koeln.de
pubmed:publicationType
Journal Article, English Abstract, Case Reports