pubmed:abstractText |
This study evaluated spatial Phi(th) inside a 70kg water phantom using the NAA method. Fifty indium foils were placed inside the water phantom and exposed under 15MV LINAC for 2.5min to yield the 10Gy X-ray dose. The Phi(th) value at the isocenter of the water was 1.03x10(6)ncm(-2)/Gy-X, and the maximum quantity of Phi(th) appeared at the water surface along the z-axis, 3.99x10(6)ncm(-2)/Gy-X. The thermal neutron dose at isocenter of the water phantom occupied approximately 0.151% of the whole photo and neutron dose.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Sep
|
pubmed:issn |
1872-9800
|
pubmed:author | |
pubmed:issnType |
Electronic
|
pubmed:volume |
68
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1816-21
|
pubmed:meshHeading |
pubmed-meshheading:20395153-Algorithms,
pubmed-meshheading:20395153-Neutron Activation Analysis,
pubmed-meshheading:20395153-Neutrons,
pubmed-meshheading:20395153-Phantoms, Imaging,
pubmed-meshheading:20395153-Photons,
pubmed-meshheading:20395153-Radiation Dosage,
pubmed-meshheading:20395153-Radiometry,
pubmed-meshheading:20395153-Scattering, Radiation,
pubmed-meshheading:20395153-Water
|
pubmed:year |
2010
|
pubmed:articleTitle |
Distribution of spatial photoneutrons inside a 70kg water phantom via neutron activation analysis.
|