Artificial heart valves comprising metal can cause significant MR image artifacts due to the material (susceptibility artifact) and/or electromagnetic characteristics (RF artifact). The purpose of our study was to examine current commercially available heart valve prostheses, integrate resonant circuits and to produce a prototype self-expanding heart valve for MRI-guided placement. Different types of commercially available heart valves were tested in MRI. Freshly excised porcine heart valves were sutured with 6-0 prolene into a 21 mm Nitinol stent. Resonant circuits were integrated into the heart valves and tuned to the Larmor frequency of the MRI (42.58 MHz for 1.0 T and about 64 MHz for 1.5 T and 128 MHz for 3T). The artifacts caused by the non-ferromagnetic heart valves and the Nitinol stent could be overcome. MRI signal could be enhanced using low flip angles <40 and visualization improved in 1T, 1.5T and 3T MRI. MRI-guided implantation was facilitated. A resonant circuit tuned to the Larmor frequency of the MR tomography can overcome the RF artifacts and thus improve the visualization of prosthetic heart valves.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:issn |
1365-2931
|
pubmed:author | |
pubmed:issnType |
Electronic
|
pubmed:volume |
18
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
149-55
|
pubmed:meshHeading |
pubmed-meshheading:20334499-Alloys,
pubmed-meshheading:20334499-Animals,
pubmed-meshheading:20334499-Female,
pubmed-meshheading:20334499-Heart Valve Prosthesis,
pubmed-meshheading:20334499-Heart Valve Prosthesis Implantation,
pubmed-meshheading:20334499-Magnetic Resonance Imaging,
pubmed-meshheading:20334499-Prosthesis Design,
pubmed-meshheading:20334499-Stents,
pubmed-meshheading:20334499-Swine,
pubmed-meshheading:20334499-Swine, Miniature
|
pubmed:year |
2009
|
pubmed:articleTitle |
Experimental MRI visible resonant prosthetic heart valves.
|