Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
12
pubmed:dateCreated
2009-11-25
pubmed:abstractText
We describe an unusual case of near fatal inappropriate implantable cardioverter defibrillator therapies due to atrial oversensing by a newly implanted ventricular lead. Chest X-ray revealed dislodgement of the active fixation lead to the tricuspid annulus area explaining the atrial oversensing and intermittent ventricular therapies.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
1532-2092
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
11
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1716-8
pubmed:meshHeading
pubmed:year
2009
pubmed:articleTitle
A life-threatening arrhythmia induced by inappropriate activation of an implantable cardioverter defibrillator.
pubmed:affiliation
Department of Cardiology, Rabin Medical Center, Beilinson Hospital-Petah Tikva, affiliated to the Sackler School of Medicine, Tel Aviv University, 39 Jabotinski Street, Petah Tikva, Israel.
pubmed:publicationType
Journal Article, Case Reports