pubmed:abstractText
Dynamic chest radiography using a flat panel detector (FPD) with a large field of view is expected to be a useful pulmonary functional evaluation method based on the respiratory changes in pixel value. For clinical use as a follow-up and therapeutic evaluation tool, the system must have a high degree of reproducibility in measurements of pixel values. The present study was performed to investigate the reproducibility of respiratory changes in pixel values. Dynamic chest radiographs of five normal subjects and one patient were obtained. Imaging was performed twice in each subject. The slope (X-ray translucency variation) was then calculated from the changes in pixel value from distance lung apex-diaphragm, and the slopes of two sequences were compared. The results showed there were no significant differences in changes in pixel value between the two sequences in all normal subject (5 males, p>0.05). The results indicated that the present method has reproducibility for measuring pulmonary function and also has potential as a tool for follow-up and therapeutic evaluation.
pubmed:language
jpn
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0369-4305
pubmed:author
pubmed:issnType
Print
pubmed:day
20
pubmed:volume
65
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
738-44
pubmed:dateRevised
2011-7-28
pubmed:meshHeading
pubmed:year
2009
pubmed:articleTitle
[Reproducibility of dynamic chest radiography with a flat-panel detector - respiratory changes in pixel value].
pubmed:affiliation
Division of Health Sciences, Radiology, Graduate School of Medical Science, Kanazawa University.