pubmed:abstractText
Electromagnetic fields may interfere with normal pacemaker function. Despite new device designs and bipolar leads, electromagnetic interference (EMI) remains a concern when pacemaker recipients are exposed to various household appliances. We report the observation of EMI by an induction heating (IH) rice cooker in a patient with sick sinus syndrome who was the recipient of a bipolar dual chamber-pacing system. Stored electrograms revealed episodes of inappropriate ventricular pacing, all coinciding with the opening of an IH rice cooker. Recipients of implantable medical devices must be warned to handle IH rice cookers with caution.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
1349-2365
pubmed:author
pubmed:issnType
Print
pubmed:volume
50
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
133-7
pubmed:dateRevised
2010-11-18
pubmed:meshHeading
pubmed:year
2009
pubmed:articleTitle
Electromagnetic interference with a bipolar pacemaker by an induction heating (IH) rice cooker.
pubmed:affiliation
The Second Department of Internal Medicine, University of Occupational and Environmental Health, Japan.
pubmed:publicationType
Journal Article, Case Reports