pubmed:abstractText |
The cavity-gas calibration factor Ngas for Memorial parallel-plate chambers was determined against calibrated Farmer chambers using two different calibration methods as suggested in the TG 21 protocol. These methods are: (a) 60Co calibration at 5-cm depth; and (b) high-energy electron beam calibration. The Ngas values determined from the 60Co in-phantom calibrations were approximately 1.9% higher than the values determined from the high-energy electron beams.
|
pubmed:grant | |
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:issn |
0094-2405
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
18
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
749-52
|
pubmed:dateRevised |
2007-11-14
|
pubmed:meshHeading |
pubmed-meshheading:1921881-Cobalt Radioisotopes,
pubmed-meshheading:1921881-Humans,
pubmed-meshheading:1921881-Mathematics,
pubmed-meshheading:1921881-Models, Anatomic,
pubmed-meshheading:1921881-Radiotherapy,
pubmed-meshheading:1921881-Radiotherapy, High-Energy,
pubmed-meshheading:1921881-Radiotherapy Dosage
|
pubmed:articleTitle |
Ngas values of the Memorial parallel-plate chambers determined in 60Co and high-energy electron beams.
|