pubmed:abstractText
We present a method to incorporate geometrical uncertainties into dose-volume histogram evaluation: the dose-volume population histogram (DVPH). For each dose-volume point, the probability of the plan DVH meeting the constraint is calculated. The use of DVPH for the target shows that the minimum dose to the PTV might not be representative of the minimum dose to the CTV considering geometrical uncertainties when the PTV extends into the build-up region. For OARs, the DVH obtained from DVPH with 90% confidence level is found to be different to the planning organ at risk volume (PRV) DVH recommended by ICRU 62, especially for parallel organs.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Feb
pubmed:issn
0031-9155
pubmed:author
pubmed:issnType
Print
pubmed:day
21
pubmed:volume
54
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
935-47
pubmed:meshHeading
pubmed:year
2009
pubmed:articleTitle
Dose-volume population histogram: a new tool for evaluating plans whilst considering geometrical uncertainties.
pubmed:affiliation
Medical Physics Department, Addenbrooke's Hospital, Cambridge CB2 0QQ, UK.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't