Source:http://linkedlifedata.com/resource/pubmed/id/19107178
Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
1
|
pubmed:dateCreated |
2008-12-24
|
pubmed:abstractText |
We present a measurement uncertainty evaluation for emission color and luminance of displays. The calibration procedures and the measurement uncertainties of a CCD-based spectroradiometer and a filter-type luminance meter are discussed. As evaluation examples, a Commission Internationale de l' Eclairage illuminant A, a liquid-crystal display with a light-emitting diode backlight, and a liquid-crystal display with a cold-cathode fluorescent lamp backlight are evaluated. The uncertainties in Commission Internationale de l' Eclairage 1931 (x,y) chromaticity coordinates are determined to be in a range from 0.0020 to 0.0050 (k=2). The uncertainties in luminance are obtained from 4.8% to 8.5% by using the spectroradiometer, while the filter-type luminance meter shows the uncertainties from 1.1% to 1.6% (k=2).
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:status |
PubMed-not-MEDLINE
|
pubmed:month |
Jan
|
pubmed:issn |
1539-4522
|
pubmed:author | |
pubmed:issnType |
Electronic
|
pubmed:day |
1
|
pubmed:volume |
48
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
99-105
|
pubmed:year |
2009
|
pubmed:articleTitle |
Measurement uncertainty evaluation for emission color and luminance of displays.
|
pubmed:affiliation |
Division of Physical Metrology, Korea Research Institute of Standards and Science, Daejeon 305-340, Korea. jhwang@kriss.re.kr
|
pubmed:publicationType |
Journal Article
|