pubmed:abstractText
A critically ill patient had a major mitral paravalvular leak, failed device closure of the leak, and severe hemolytic anemia after a third mitral valve replacement. Echocardiography showed the Amplatzer occluder device completely displaced and swirled in the left atrial cavity. His condition was greatly improved after a fourth mitral valve replacement.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
1569-9285
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
7
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1131-3
pubmed:meshHeading
pubmed-meshheading:18753152-Aged, pubmed-meshheading:18753152-Anemia, Hemolytic, pubmed-meshheading:18753152-Echocardiography, Doppler, Color, pubmed-meshheading:18753152-Echocardiography, Transesophageal, pubmed-meshheading:18753152-Equipment Design, pubmed-meshheading:18753152-Equipment Failure, pubmed-meshheading:18753152-Foreign-Body Migration, pubmed-meshheading:18753152-Heart Catheterization, pubmed-meshheading:18753152-Heart Valve Diseases, pubmed-meshheading:18753152-Heart Valve Prosthesis, pubmed-meshheading:18753152-Heart Valve Prosthesis Implantation, pubmed-meshheading:18753152-Humans, pubmed-meshheading:18753152-Male, pubmed-meshheading:18753152-Mitral Valve, pubmed-meshheading:18753152-Prosthesis Failure, pubmed-meshheading:18753152-Recurrence, pubmed-meshheading:18753152-Reoperation, pubmed-meshheading:18753152-Treatment Outcome
pubmed:year
2008
pubmed:articleTitle
Displacement of the Amplatzer occluder device from the mitral paraprosthetic leak.