pubmed:abstractText
Evaluation of luminal temperature during left atrial ablation is used clinically; however, luminal temperature does not necessarily reflect temperature within the esophageal wall and poses a risk of atrioesophageal fistula. This animal study evaluates luminal esophageal temperature and its relation to the temperature of the external esophageal tissue during left atrial lesions using the 8 mm solid tip and the open irrigated tip catheters (OIC).
pubmed:commentsCorrections
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
1540-8167
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
19
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
641-4
pubmed:dateRevised
2010-11-18
pubmed:meshHeading
pubmed:year
2008
pubmed:articleTitle
Esophageal luminal temperature measurement underestimates esophageal tissue temperature during radiofrequency ablation within the canine left atrium: comparison between 8 mm tip and open irrigation catheters.